| Title |
Beam Quality Assurance for Proton Clinical Beams at MedAustron |
| Authors |
- L.C. Penescu, F. Farinon, A. Garonna, M. Kronberger, T.K.D. Kulenkampff, C. Kurfürst, S. Myalski, S. Nowak, F. Osmić, M.T.F. Pivi, C. Schmitzerpresenter, P. Urschütz, A. Wastl
EBG MedAustron, Wr. Neustadt, Austria
|
| Abstract |
The commissioning process of the MedAustron accelerator has delivered the configurations providing the requested beam parameters in the irradiation room, and at the same time it identified the critical points where a performance drift can appear. The strategy for beam quality assurance has therefore two components: testing the specific parameters of the beam delivered to the irradiation room, and testing for any drifts that might appear at the critical points. We present here the monitoring strategy, the observed limitations, the tools employed and the long-term statistics of the beam quality assurance for proton clinical beams.
|
| Paper |
download TUPOY001.PDF [0.610 MB / 3 pages] |
| Export |
download ※ BibTeX
※ LaTeX
※ Text/Word
※ RIS
※ EndNote |
| Conference |
IPAC2016, Busan, Korea |
| Series |
International Particle Accelerator Conference (7th) |
| Proceedings |
Link to full IPAC2016 Proccedings |
| Session |
Poster Session |
| Date |
10-May-16 16:00–18:00 |
| Main Classification |
08 Applications of Accelerators |
| Sub Classification |
U01 Medical Applications |
| Keywords |
operation, proton, controls, hardware, synchrotron |
| Publisher |
JACoW, Geneva, Switzerland |
| Editors |
Christine Petit-Jean-Genaz (CERN, Geneva, Switzerland); Dong Eon Kim (PAL, Pohang, Republic of Korea); Kyung Sook Kim (PAL, Pohang, Republic of Korea); In Soo Ko (POSTECH, Pohang, Republic of Korea); Volker RW Schaa (GSI, Darmstadt, Germany) |
| ISBN |
978-3-95450-147-2 |
| Published |
June 2016 |
| Copyright |
|